48-7009-E 7-9
7.2 Accessories
7.2.1 Current transformer
Current transformer is used to measure current values of a heater through connecting it to TOHO’s
digital controller TTM-200 series and the heater.
Model: CTL-6-P-H-(S)
Product specifications
Current applicable
0.1 - 80.0 A rms (50/60 Hz)
* This is specification for CT main frame.
Max. allowable current
80 A rms continuous
Withstand voltage
2000 VAC for a minute (between through-hole and output terminal)
Isolation resistance
500 VDC, 100 M or larger (between through-hole and output terminal)
Use temperature range
-20 to 70 °C
Storage temperature range
-30 to 90 °C
Weight
Approx. 16 g
Accessories
None
Outside dimension
7.2.2 Loader cable
Loader cable is used to connect TOHO’s digital controller TTM-200 series and a personal computer
(PC) and such so that various values are set and read on the PC, using the dedicated loader software
installed on it.
Model: TTM-LOADER
Product specifications
OS to meet
Windows XP
USB I/F standards
In compliance with USB Specification 2.0
DTE speed
Up to 38400 bps
Connector specification
PC side: USB
Controller side: 2.5 mm stereo jack
Consumption current
90 mA
Use temperature range
0 to 50 °C
Use humidity range
20 to 90 %RH (no dew condensation allowed)
Storage temperature range
-10 to 60 °C (no freeze or dew condensation allowed)
Storage humidity range
5 to 95 %RH (no dew condensation allowed)
Power supply voltage
Bus-powered (to be supplied from PC side)
Weight
40 g
Outside dimension
USB connector