TOHO TOHO ELECTRONICS INC.
DWG. No.
45-4199-E
PAGE
59/66
4.4.13 Setting mode for CH 1 polygonal line approximation
Polygonal line approximation is for outputting
when inputting
.
Character
PV screen
Auxiliary screen
Designation Description Initial value
1
□
□□□□
Select CH 1
polygonal line
approximation
Setting concerning polygonal line approximation for
CH 1
2
to
17
to
Set an input for
polygonal line
approximation 0 to
F
Thermocouple input
Setting range: -199 to 999 or -199.9 to 999.9
Setting unit: °C
Resistance bulb input
Setting range: -199 to 999 or -199.9 to 999.9 or
-199.99 to 999.99
(when the input type is
,
,
or
)
Setting unit: °C
Current and voltage inputs
Setting range: -19999 to 99999
The decimal place can be set to a
desired position.
Setting unit: digit
The setting is limited so that polygonal line
approximation n
≦
polygonal line approximation n + 1.
_PAA0 is
_PAA1 to F is
CH2 only
Potentiometer input
Setting range: -199.9 to 999.9
Setting unit: %
18
to
33
to
Set an output for
polygonal line
approximation 0 to
F
Thermocouple input
Setting range: -199 to 999 or -199.9 to 999.9
Setting unit: °C
Resistance bulb input
Setting range: -199 to 999 or -199.9 to 999.9 or
-199.99 to 999.99
(when the input type is
,
,
or
)
Setting unit: °C
Current and voltage inputs
Setting range: -19999 to 99999
The decimal place can be set to a
desired position.
Setting unit: digit
The setting is limited so that polygonal line
approximation n
≦
polygonal line approximation n + 1.
_PAB0 is
_PAB1 to F is
CH2 only
Potentiometer input
Setting range: -199.9 to 999.9
Setting unit: %
Kommentare zu diesen Handbüchern